7 mm C.2 Download CK_7MMC2

 

 

 

 

 

 

 

 

 

 

 

 

 

Standard Definitions

 

 

Standard Type C0/L0 Term Z C1/L1 C2/L2 C3/L3 Load Type Offset Delay Offset Z0 Offset Loss Min Freq Max Freq Medium Standard Label
#1 Short 0.3566 -33.39201 1.7542 -0.0336 Fixed 0 50 0 0 999 Coax  SHORT
#2 Open 90.47998 763.59985 -63.81801 6.4337 Fixed 0 50 0 0 999 Coax  OPEN
#3 Short 0 0 0 0 Fixed 0 50 0 0 999 Coax  
#4 Short 0 0 0 0 Fixed 0 50 0 0 999 Coax  
#5 Short 0 0 0 0 Fixed 0 50 0 0 999 Coax  
#6 Short 0 0 0 0 Fixed 0 50 0 0 999 Coax  
#7 Short 0 0 0 0 Fixed 0 50 0 0 999 Coax  
#8 Short 0 0 0 0 Fixed 0 50 0 0 999 Coax  
#9 Load

 

Fixed 0 50 0 0 999 Coax BROADBAND
#10 Load

 

Fixed 0 50 0 0 999 Coax  
#11 Thru

 

Fixed 0 50 0 0 999 Coax  THRU
#12 Load

 

Fixed 0 50 0 0 2.001 Coax LOWBAND
#13 Thru

 

Fixed 0 50 0 0 999 Coax undefined
#14 Thru

 

Fixed 0 50 0 0 999 Coax THRU
#15 Thru

 

Fixed 23.18999 49.988 0.7 0 999 Coax 2-18 LINE
#16 Thru

 

Fixed 1E+12 50 0.7 0 2.001 Coax 0-2 LOADS
#17 Short 0 0 0 0 Fixed 0 50 0 0 999 Coax  
#18 Short 0 0 0 0 Fixed 0 50 0 0 999 Coax  
#19 Short 0 0 0 0 Fixed 0 50 0 0 999 Coax  
#20 Load

 

Offset 23.18999 49.988 0.7 1.999 999 Coax OFFSET
#21 Short 0 0 0 0 Fixed 23.18999 49.988 0.7 1.999 999 Coax OFFSET

 

Class Assignments

 

 

A B C D E F G Standard Class Label

 

S11A 2 21

 

OPENS

 

S11B 1

 

SHORT

 

S11C 9 12 20

 

LOADS

 

S22A 2 21

 

OPENS

 

S22B 1

 

SHORT

 

S22C 9 12 20

 

LOADS

 

Forward Transmission 11

 

THRU

 

Reverse Transmission 11

 

THRU

 

Forward Match 11

 

THRU

 

Reverse Match 11

 

THRU

 

Forward Isolation 9

 

ISOL'N STD

 

Reverse Isolation 9

 

ISOL'N STD

 

Frequency Response 1 2 11

 

RESPONSE

 

TRL Thru or Line 14

 

THRU

 

TRL Reflect 1

 

SHORT

 

TRL Line or Match 15 16

 

LINE

 

Adapter 13

 

undefined

 

 

VNA = Agilent 8510C

 

System Z0 = 50 ohm

 

TRL Set Ref = Thru

 

TRL Cal Z0 = Line Z0

 

TRL Lowband Frequency = 2 GHz