N-50 Download D53N5F.zip
Standard Definitions
Standard Type C0/L0 Term Z C1/L1 C2/L2 C3/L3 Load Type Offset Delay Offset Z0 Offset Loss Min Freq Max Freq Medium Standard Label
#1 Short 0 0 0 0 Fixed 45.955 50 1.087 0 999 Coax  SHORT
#2 Open 89.939 2536.8 -264.99 13.4 Fixed 41.03 50 0.93 0 999 Coax  OPEN 
#3 Load Fixed 0 50 2.3 0 999 Coax BROADBAND
#4 Thru Fixed 0 50 2.3 0 999 Coax  THRU 
#5 Load Sliding 0 50 1.3 3.999 999 Coax SLIDING
#6 Load Fixed 0 50 1.3 0 4.001 Coax LOWBAND
#7 Short 0 0 0 0 Fixed 45.955 50 1.087 0 999 Coax  SHORT
#8 Open 89.939 2536.8 -264.99 13.4 Fixed 40.74 50 0.93 0 999 Coax  OPEN 
Class Assignments
A B C D E F G Standard Class Label
S11A 2 OPEN 
S11B 1 SHORT
S11C 3 LOAD 
S22A 2 OPEN 
S22B 1 SHORT
S22C 3 LOAD 
Forward Transmission 4 THRU
Reverse Transmission 4 THRU
Forward Match 4 THRU
Reverse Match 4 THRU
Response 1 2 4 RESPONSE
Response & Isolation 1 2 4 RESPONSE
TRL Thru or Line 4 THRU
TRL Reflect 2 OPEN 
TRL Line or Match 3 LOAD 
VNA = Agilent 8719D/E
System Z0 = 50 ohm
TRL Set Ref = Thru
TRL Cal Z0 = Line Z0
TRL Lowband Frequency = 2 GHz